Phlebology 2008;23:69-76
doi:10.1258/phleb.2007.007038
© 2008 Royal Society of Medicine Press
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Google Scholar
Right arrow Articles by Disselhoff, B C V M
Right arrow Articles by Moll, F L
PubMed
Right arrow PubMed Citation
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Original article

Endovenous laser ablation: an experimental study on the mechanism of action

B C V M Disselhoff*, A I Rem{dagger}, R M Verdaasdonk{dagger}, D J der Kinderen{ddagger} and F L Moll§

* Department of Surgery, Mesos Medical Centre; {dagger} Department of Clinical Physics, University Medical Centre Utrecht; {ddagger} Department of Dermatology, Mesos Medical Centre; § Department of Vascular Surgery, University Medical Centre Utrecht, Utrecht, The Netherlands

Correspondence: B C V M Disselhoff MD, Department of Surgery, Mesos Medical Centre, PO Box 8605, 3503 RP Utrecht, The Netherlands. Email: bcvmdisselhoff{at}mesos.nl

Objectives: The aim of this experimental study was to investigate the mechanism of action of endovenous laser ablation (EVLA) using an 810-nm diode laser.

Methods: We compared intermittent and continuous delivery of laser energy and studied the absorption of laser light by blood, intravascular temperatures in ex vivo human vein segments using an intravascular thermography catheter and heat dissipation in a model tissue using the Schlieren technique.

Results: Laser light is absorbed by blood and converted to heat leading to coagulation, vaporization and carbonization, and forming an isolating layer at the fibre tip. Laser energy is then absorbed into the isolating layer forming black patches that burned on the laser fibre. Intravascular temperature increased rapidly above carbonization temperatures (300°C) after the fibre tip reached the thermocouple, stayed at this temperature for a few seconds and decreased gradually to around 30°C, 10 s after the fibre tip passed the thermocouple. Schlieren techniques revealed that heat spread from the laser was locally distributed and closely around the laser fibre tip while heat dissipation is minimal and comparable for both exposures. Compared with intermittent exposure, continuous exposure results in more carbonization, higher mean maximum intravascular temperature (128 ± 7 vs. 75 ± 4°C), and long-lasting temperature of 100°C (1.2 ± 0.4 vs. 0.1 ± 0.1 s).

Conclusion: In this experimental study, application of endovenous laser shows to be dominated by carbonization at the fibre tip. Although intraluminal laser-induced heat was heterogeneously distributed, with laser tip temperatures up to 1200°C, heat dissipation was minimal. Continuous exposure of laser light appears to be better suited in EVLA than intermittent.

Key Words: endovenous laser ablation • blood absorption of laser light • intravascular temperature measurements • heat dissipation